The Fluke 1587 FC Insulation Multimeter features four powerful new diagnostic functions through the Fluke Connect Measurements app:

  • Pl/DAR timed ratio tests with TrendIt™ graphs identifies moisture and contaminated insulation problems faster
  • Memory storage through Fluke Connect eliminates writing down results, reduces errors and saves data for historical tracking over time
  • Temperature Compensation through app for establishing accurate baselines and relevant historical comparisons
  • Historical tracking and trending of assets identifies degradation over time, allows real-time decisions to be made in the field with Fluke Connect® Assets

The Fluke i400 AC Current Clamp is a high-safety current clamp for digital multimeters. The current clamp provides accurate current measurements without breaking electric circuits, helping you take measurements safely with industry standard safety ratings.

Fluke 9040 Phase Rotation Indicator is effective for measuring phase rotation in all areas where three phase supplies are used to feed motors, drives and electrical systems. It allows rapid determination of phase sequence and ideally suited for commercial and industrial applications.

Electrical Specifications

  • AC voltage measurement
    • Range: 600.0 mV
      • Resolution: 0.1 mV
      • Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts): ±(1% + 3)
      • Accuracy 60 Hz to 5000 Hz ±(% of Rdg + Counts): ±(2% + 3)
    • Range: 6.000 V
      • Resolution: 0.001 V
      • Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts): ±(1% + 3)
      • Accuracy 60 Hz to 5000 Hz ±(% of Rdg + Counts): ±(2% + 3)
    • Range: 60.00 V
      • Resolution: 0.01 V
      • Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts): ±(1% + 3)
      • Accuracy 60 Hz to 5000 Hz ±(% of Rdg + Counts): ±(2% + 3)
    • Range: 600.0 V
      • Resolution: 0.1 V
      • Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts): ±(1% + 3)
      • Accuracy 60 Hz to 5000 Hz ±(% of Rdg + Counts): ±(2% + 3)¹
    • Range: 1000 V
      • Resolution: 1 V
      • Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts): ±(2% + 3)
      • Accuracy 60 Hz to 5000 Hz ±(% of Rdg + Counts): ±(2% + 3)¹

¹ 1 kHz bandwidth

Low-Pass Filter Voltage 

  • Range: 600.0 mV
    • Resolution: 0.1 mV
    • Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts): ±(1% + 3)
    • Accuracy 60 Hz to 400 Hz ±(% of Rdg + Counts): +(2% + 3), -(6% - 3)
  • Range: 6.000 V
    • Resolution: 0.001 V
    • Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts): ±(1% + 3)
    • Accuracy 60 Hz to 400 Hz ±(% of Rdg + Counts): +(2% + 3), -(6% - 3)
  • Range: 60.00 V
    • Resolution 0.01 V
    • Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts): ±(1% + 3)
    • Accuracy 60 Hz to 400 Hz ±(% of Rdg + Counts): +(2% + 3), -(6% - 3)
  • Range: 600.0 V
    • Resolution: 0.1 V
    • Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts): ±(1% + 3)
    • Accuracy 60 Hz to 400 Hz ±(% of Rdg + Counts): +(2% + 3), -(6% - 3)
  • Range: 1000 V
    • Resolution: 1 V
    • Accuracy 50 Hz to 60 Hz ±(% of Rdg + Counts): ±(2% + 3)
    • Accuracy 60 Hz to 400 Hz ±(% of Rdg + Counts): +(2% + 3), -(6% - 3)

DC Voltage Measurement

  • Range: 6.000 V dc
    • Resolution: 0.001 V
    • Accuracy ±(% of Rdg + Counts): ±(0.09% + 2)
  • Range: 60.00 V dc
    • Resolution: 0.01 V
    • Accuracy ±(% of Rdg + Counts): ±(0.09% + 2)
  • Range: 600.0 V dc
    • Resolution: 0.1 V
    • Accuracy ±(% of Rdg + Counts): ±(0.09% + 2)
  • Input impedance: 10 MΩ (nominal), <100 pF
  • Normal mode rejection ratio: >60 dB @ 50 Hz or 60 Hz
  • Common mode rejection ratio: >120 dB @ dc, 50 Hz or 60 Hz (1 k unbalance)
  • Accuracies apply to ± 100% of range

DC Millivolts Measurement

  • Range: 600.0 mVdc
    • Resolution: 0.1 mV
    • Accuracy ±(% of Rdg + Counts): ±(0.1% + 1)

DC and AC Current Measurement

  • AC 45 Hz to 1000 Hz
    • Range: 400 mA
    • Resolution: 0.1 mA
    • Accuracy ±(% of Rdg+Counts): ±(1.5% + 2)¹
    • Burden voltage (Typical): 2 mV/mA
    • Range: 60 mA 
    • Resolution: 0.01 mA 
    • Accuracy ±(% of Rdg+Counts): ±(1.5% + 2)¹
    • Burden voltage (Typical): 2 mV/mA

DC

  • Range: 400 mA
  • Resolution: 0.1 mA
  • Accuracy ±(% of Rdg+Counts): ±(0.2% + 2)
  • Burden voltage (Typical): 2 mV/mA
  • Range: 60 mA
  • Resolution: 0.01 mA
  • Accuracy ±(% of Rdg+Counts): ±(0.2% + 2)
  • Burden voltage (Typical): 2 mV/mA
  • Overload: 600 mA for 2 minutes maximum
  • Fuse protection for mA input: 0.44 mA, 1000 V, IR 10 kA
  • AC conversion: Inputs are ac-coupled and calibrated to the rms value of sine wave input

Conversions are true-rms responding and specified from 5% to 100% of range. Input signal crest factor can be up to 3 up to 300 mA, decreasing linearly to crest factor ≤1.5 at 600 mA. For non-sinusoidal waveforms add +(2% reading + 2% FS) typical, for a crest factor up to 3.

¹ 1 kHz bandwidth

Ohms Measurement

  • Range: 600.0 Ω
    • Resolution: 0.1 Ω
    • Accuracy +(% of Rdg+Counts)¹: ±(0.9% + 2)
  • Range: 6.000 kΩ
    • Resolution: 0.001 kΩ
    • Accuracy +(% of Rdg+Counts)¹: ±(0.9% + 2)
  • Range: 60.00 kΩ
    • Resolution: 0.01 kΩ 
    • Accuracy +(% of Rdg+Counts)¹: ±(0.9% + 2)
  • Range: 600.0 kΩ
    • Resolution: 0.1 KΩ 
    • Accuracy +(% of Rdg+Counts)¹: ±(0.9% + 2)
  • Range: 6.000 MΩ
    • Resolution: 0.001 MΩ 
    • Accuracy +(% of Rdg+Counts)¹: ±(0.9% + 2)
  • Range: 50.0 MΩ ²
    • Resolution: 0.01 MΩ 
    • Accuracy +(% of Rdg+Counts)¹: ±(1.5% + 3)
  • Overload protection: 1000 V rms or dc
  • Open circuit test voltage: <8.0 V dc
  • Short circuit current: <1.1 mA

¹ Accuracies apply from 0% to 100% of range

² Up to 80% relative humidity

Diode Test

  • Diode test indication: Display voltage drop: 0.6 V at1.0 mA nominal test current:
  • Accuracy: ±(2% + 3)

Continuity Test

  • Continuity indication: Continuous audible tone for testresistance below 25 Ω and off above 100 Ω. Maximum Reading; 1000 Ω
  • Open circuit voltage: <8.0 V
  • Short circuit current: 1.0 mA typical
  • Overload protection: 1000 V rms
  • Response time: >1 m sec

Frequency Measurement

  • Range: 99.99 Hz 
  • Resolution: 0.01 Hz 
  • Accuracy +(% of Rdg+Counts)¹: ±(0.1% + 1)
  • Range: 999.9 Hz 
  • Resolution: 0.1 Hz 
  • Accuracy +(% of Rdg+Counts)¹: ±(0.1% + 1)
  • Range: 9.999 kHz 
  • Resolution: 0.001 kHz 
  • Accuracy +(% of Rdg+Counts)¹: ±(0.1% + 1)
  • Range: 99.99 kHz 
  • Resolution: 0.01 kHz 
  • Accuracy +(% of Rdg+Counts)¹: ±(0.1% + 1)

Frequency Counter Sensitivity

  • 600.0 mV ac:
    • V ac Sensitivity (RMS Sine Wave)¹ 5 Hz to 20 kHz: 100.0 mV 
    • V ac Sensitivity (RMS Sine Wave)¹ 20 kHz to 100 kHz: 150.0 mV 
    • DC Trigger Levels¹ to 20 kHz²: NA
  • 6.0 V: 
    • V ac Sensitivity (RMS Sine Wave)¹ 5 Hz to 20 kHz: 1.0 V 
    • V ac Sensitivity (RMS Sine Wave)¹ 20 kHz to 100 kHz: 1.5 V 
    • DC Trigger Levels¹ to 20 kHz²: -400.0 mV and 2.5 V
  • 60.0 V:
    • V ac Sensitivity (RMS Sine Wave)¹ 5 Hz to 20 kHz: 10.0 V 
    • V ac Sensitivity (RMS Sine Wave)¹ 20 kHz to 100 kHz: 36.0 V 
    • DC Trigger Levels¹ to 20 kHz²: 1.2 V and 4.0 V
  • 600.0 V:
    • V ac Sensitivity (RMS Sine Wave)¹ 5 Hz to 20 kHz: 100.0 V 
    • V ac Sensitivity (RMS Sine Wave)¹ 20 kHz to 100 kHz: -
    • DC Trigger Levels¹ to 20 kHz²: 12.0 V and 40.0 V
  • 1000.0 V:
    • V ac Sensitivity (RMS Sine Wave)¹ 5 Hz to 20 kHz: 300.0 V 
    • V ac Sensitivity (RMS Sine Wave)¹ 20 kHz to 100 kHz: -
    • DC Trigger Levels¹ to 20 kHz²: 12.0 V and 40.0 V

¹ Maximum input for specified accuracy = 10x range (1000 V max). Noise at low frequencies and amplitudes may affect accuracy

² Usable to 100 kHz with full scale input

Capacitance

  • Range: 1000 nF 
    • Resolution: 1 nF
    • Accuracy ±(% of Rdg+Counts): ±(1.2% + 2)
  • Range: 10.00 μF 
    • Resolution: 0.01 μF
    • Accuracy ±(% of Rdg+Counts): ±(1.2% + 2)
  • Range: 100.0 μF 
    • Resolution: 0.1 μF
    • Accuracy ±(% of Rdg+Counts): ±(1.2% ±90 counts)
  • Range: 9999 μF
    • Resolution: 1 μF
    • Accuracy ±(% of Rdg+Counts): (1.2% ±90 counts)

Temperature Measurement

  • Range: -40 ° C to 537 ° C 
    • Resolution: 0.1 °C 
    • Accuracy ±(% of Rdg+Counts): ±(1% + 10 counts)
  • Range: -40 ° F to 998 ° F 
    • Resolution: 0.1 °F 
    • Accuracy ±(% of Rdg+Counts): ±(1% + 18 counts)

¹ Accuracies apply following 90 minutes settling time after a change in the ambient temperature of the instrument

Insulation Specifications

  • Measurement range: 0.01 MΩ to 2 GΩ
  • Test voltages: 50, 100, 250, 500, 1000 V
  • Test voltage accuracy: +20%, -0%
  • Short-circuit test current: 1 mA nominal
  • Auto discharge: Discharge time <0.5 second for C = 1 μF or less
  • Live circuit detection: Inhibit test if terminal voltage > 30 V prior to initialization of test
  • Maximum capacitive load: Operable with up to 1 μF load

Output Voltage

  • 50 V (0% to +20%): 
    • Display range: 0.01 to 6.00 MΩ
    • Resolution: 0.01 MΩ
    • Test current: 1 mA @ 50 kΩ
    • Resistance accuracy ±(% of Rdg + Counts): ±(3% + 5 counts)
    • Display range: 6.0 to 50.0 MΩ
    • Resolution: 0.1 MΩ
    • Test current: 1 mA @ 50 kΩ
    • Resistance accuracy ±(% of Rdg + Counts): ±(3% + 5 counts)
  • 100 V (0% to +20%): 
    • Display range: 0.01 to 6.00 MΩ
    • Resolution: 0.01 MΩ
    • Test current: 1 mA @ 100 kΩ
    • Resistance accuracy ±(% of Rdg + Counts): ±(3% + 5 counts)
    • Display range: 6.0 to 60.0 MΩ
    • Resolution: 0.1 MΩ
    • Test current: 1 mA @ 100 kΩ
    • Resistance accuracy ±(% of Rdg + Counts: ±(3% + 5 counts)
    • Display range: 60 to 100 MΩ 
    • Resolution: 1 MΩ
    • Test current: 1 mA @ 100 kΩ
    • Resistance accuracy ±(% of Rdg + Counts): ±(3% + 5 counts)
  • 250 V (0% to +20%):
    • Display range: 0.1 to 60.0 MΩ 
    • Resolution: 0.1 MΩ
    • Test current: 1 mA @ 250 kΩ 
    • Resistance accuracy ±(% of Rdg + Counts): ±(1.5% + 5 counts)
    • Display range: 60 to 250 MΩ 
    • Resolution: 1 MΩ
    • Test current: 1 mA @ 250 kΩ 
    • Resistance accuracy ±(% of Rdg + Counts): ±(1.5% + 5 counts)
  • 500 V (0% to +20%)
    • Display range: 0.1 to 60.0 MΩ
    • Resolution: 0.1 MΩ
    • Test current: 1 mA @ 500 kΩ 
    • Resistance accuracy ±(% of Rdg + Counts): ±(1.5% + 5 counts)
    • Display range: 60 to 500 MΩ
    • Resolution: 1 MΩ
    • Test current: 1 mA @ 500 kΩ 
    • Resistance accuracy ±(% of Rdg + Counts): ±(1.5% + 5 counts)
  • 1000 V (0% to +20%):
    • Display range: 0.1 to 60.0 MΩ 
    • Resolution: 0.1 MΩ
    • Test current: 1 mA @ 1 MΩ
    • Resistance accuracy ±(% of Rdg + Counts): ±(1.5% + 5 counts)
    • Display range: 60 to 600 MΩ 
    • Resolution: 1 MΩ
    • Test current: 1 mA @ 1 MΩ
    • Resistance accuracy ±(% of Rdg + Counts): ±(1.5% + 5 counts)
    • Display range: 0.6 to 2.0 GΩ 
    • Resolution: 100 MΩ 
    • Test current: 1 mA @ 1 MΩ
    • Resistance accuracy ±(% of Rdg + Counts): ±(10% + 3 counts)

General Specifications

  • Maximum voltage applied to any terminal and common: 1000 V
  • Storage temperature: -40 °C to 60 °C (-40 °F to 140 °F)
  • Operating temperature: -20 °C to 55 °C (-4 °F to 131 °F)
  • Temperature coefficient: 0.05 x (specified accuracy) per °C for temperatures <18 °C or >28 °C (<64 °F or >82 °F)
  • Relative humidity:
    • Noncondensing
    • 0% to 95% @ 10 °C to 30 °C (50 °F to 86 °F)
    • 0% to 75% @ 30 °C to 40 °C (86 °F to 104 °F)
    • 0% to 40% @ 40 °C to 55 °C (104 °F to 131 °F)
  • Vibration: Random, 2 g, 5-500 Hz per MIL-PRF-28800F, Class 2 instrument
  • Radio frequency communication: 2.4 GHz ISM Band
  • Radio frequency certification: FCC: T68-FBLE, IC: 6627A-FBLE

Electromagnetic Compatibility 

  • International:
    • IEC 61326-1:Portable Electromagnetic Environment; IEC 61326-2-2 CISPR 11: Group 1, Class A
    • Group 1: Equipment has intentionally generated and/or uses conductively-coupled radio frequency energy that is necessary for the internal function of the equipment itself.
    • Class A: Equipment is suitable for use in all establishments other than domestic and those directly connected to a low-voltage power supply network that supplies buildings used for domestic purposes. There may be potential difficulties in ensuring electromagnetic compatibility in other environments due to conducted and radiated disturbances.
    • Emissions that exceed the levels required by CISPR 11 can occur when the equipment is connected to a test object. The equipment may not meet the immunity requirements of this standard when test leads and/or test probes are connected.
  • Korea (KCC):
    • Class A Equipment (Industrial Broadcasting & Communication Equipment)
    • Class A: Equipment meets requirements for industrial electromagnetic wave equipment and the seller or user should take notice of it. This equipment is intended for use in business environments and not to be used in homes.
  • USA (FCC): 47 CFR 15 subpart B. This product is considered an exempt device per clause 15.103.
  • Enclosure protection: IEC 60529: IP40 (non-operating)

Safety

  • IEC 61010-1: Pollution Degree 2
  • IEC 61010-2-033: CAT IV 600 V / CAT III 1000 V
  • Batteries: Four AA batteries (NEDA 15A or IEC LR6)
  • Battery life: Meter use 1000 hours; Insulation test use: Meter can perform at least 1000 insulation tests with fresh alkaline batteries at room temperature. These are standard tests of 1000 V into 1 MΩ with a duty cycle of 5 seconds on and 25 seconds off.
  • Size (H x W x L): 5.0 x 10.0 x 20.3 cm (1.97 x 3.94 x 8.00 in)
  • Weight : 550 g (1.2 lb)
  • Altitude:
    • Operating: 2000 m
    • Storage: 12,000 m
  • Over-range capability: 110% of range except for capacitance which is 100%
  • Frequency overload protection: <107 V Hz
  • Fuse protection for mA input: 0.44A, 1000 V, IR 10 kA
  • Fluke i400 AC Current Clamp

Specifications

  • Nominal current range: 400 A
  • Continuous current range: 1 A - 400 A
  • Maximum non-destructive current: 1000 A
  • Lowest measurable current: 1 A
  • Basic accuracy:
    • 2% + 0.06 A
    • (45-400 Hz)
    • (% reading + floorspec)
  • Useable frequency: 5 Hz - 20 kHz
  • Output level(s): 1 mA/A

Safety Specifications

  • Safety: CAT IV 600 V, CAT III, 1000 V
  • Maximum voltage: 1000 V AC rms from earth to ground, in compliance with EN61010

Mechanical and General Specifications

  • Warranty: 1 year
  • Maximum conductor diameter: 32 mm (1.25 in)
  • Maximum conductor size (busbar): 750 MCM
  • Output cable length: 1.5 m (59 in)
  • Shrouded banana plugs: Yes
  • Fluke 9040 Phase Rotation Indicator

Technical Specifications

  • Voltage range: 40-700 V
  • Frequency range: 15-400 Hz
  • Operating timer: Continuous

Environmental Specifications

  • Operating temperature: 0 °C to 40 °C
  • Dust/water resistance: IP 40

Safety Specifications

  • Electrical safety: EN 61010, EN 61557-7
  • Overvoltage category:
  • CAT III, 600 V
  • CAT IV, 300V

Mechanical and General Specifications

  • Size: 124 x 61 x 27 mm
  • Weight: 200 g
  • Power supply: From unit under test
  • Warranty: 2 years
Specification
Part No FLU1587MDTFC
Barcode 95969797283
Brand Fluke
Warranty 2 Year
Net Weight (kg) 3.2